Visual checks

1 Radiation protection

Make sure that there is no mechanical damage on the POWERPHOS tube housing
assembly and image intensifier which could impair the radiation protection.
Damage present (brief description):

2 Checking the DHHS and Identification Labels

Confirm that the certification labels as well as serial number and model number labels are
legible. Use the illustrations which are shown in the Operating Instructions of the product.
If labels are required, contact the SIEMENS Medical Solutions Local Service Organization
or the SIEMENS Medical Solutions Regulatory Affairs department.
Component certified
ARCADIS Avantic
X-ray control
Beam Limiting device
BLD
Powerphos Avantic
Tube housing assembly
Image Intensifier
Warning label present and legible
Siemens AG
Medical Solutions
Powerphos and image intensifier
have been checked for mechanical
damage
Model No.
Serial No.
SPR2-330.662.01
Rev. 02
05.06
3
YES
NO
Date
YES