Installation
!
!
!
40
3.7.1
Shortening the connecting cable
A recalibration must be performed before commissioning see Page 56 ff.
Note!
The maximum connection length between the probe and separate housing is 6 m and is indicated
by dimension L4. When ordering a device with a separate housing, the connection length must be
specified.
If the connecting cable is to be shortened or guided through a wall, it must be disconnected from
the process connection. To do so, proceed as follows:
• Unscrew the pressing screw (1) using a 22mm open-end wrench. If necessary, hold the process
connection. In doing so, ensure that neither the connecting cable nor the probe is turned in the
process.
• Pull the insert seal (2) out of the cable gland (3).
• Using a 22mm open-end wrench, disconnect the cable gland (3) from the adapter disk. If
necessary, hold it against the adapter disk (6) using a 34mm open-end wrench.
• Disconnect the adapter disk (6) from the sleeve (17).
• Remove the snap ring using a snap ring pliers.
• Using pliers, grab the nut (M6) on the Multilam plug and pull out the plug.
• Disconnect the connection between the yellow and red (guard) wires.
• Then, shorten the connecting cable to the desired length. If the separate housing is in a different
room than the probe, you can now route the connecting cable through the wall.
• You can now reassemble the device by following the reverse order of steps.
Note!
• If you shorten the connecting cable, we recommend reusing all strands with ring terminals.
• If the wires are not reused, the crimp connections of the newly attached eyelets must be insulated,
for example using a heat-shrinking sleeve (danger of short circuit).
• All soldered joints must be insulated.
3.8

Installing bracket for wall and pipe mounting

3.8.1
Wall holder unit
Note!
The wall holder unit first has to be screwed to the separate housing before you can use it as a drilling
template. The distance between the holes is reduced by screwing it to the separate housing.
3 3
Liquicap M FTI51, FTI52
+1.1
+1.1
86
86
–0.8
–0.8
+1.2
+1.2
70
70
–0.8
–0.8
TI418F20