Installation Examples - Endress+Hauser Liquicap M FTI51 Operating Instructions Manual

Capacitive level limit switch
Hide thumbs Also See for Liquicap M FTI51:
Table of Contents

Advertisement

Liquicap M FTI51, FTI52
!
Endress+Hauser
3.4

Installation examples

3.4.1
Rod probes
Conductive tanks (metal tanks)
If the process connection of the probe is insulated from the metal tank (e.g. using seal material),
connect the ground connection on the probe housing to the tank using a short line.
Note!
• A fully insulated rod probe may be neither shortened nor extended.
• If the insulation of the probe rod is damaged, this will falsify the measurement result.
• These application examples show vertical installation for MAX level limit detection.
FTI51 rod probe
FTI51 rod probe with ground tube
Nonconductive tanks (plastic tanks)
For installation in a plastic tank, use a probe with ground tube.
Installation
L00-FTI5xxxx-11-06-xx-xx-004
L00-FTI5xxxx-11-06-xx-xx-005
31

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Liquicap m fti52

Table of Contents