Spo2 Performance Test; Nibp Performance Test; Temp Performance Test - EDAN F15 Series Service Manual

Hide thumbs Also See for F15 Series:
Table of Contents

Advertisement

F15 Series Fetal & Maternal Monitor Service Manual

10.2.4 SpO2 Performance Test

This test checks the performance of SpO
Required Equipment: SpO
Connect the equipment and the monitor as shown below:
1)
SpO
Sensor
2
Switch on the monitor.
2)
Generate the SpO2 signal 98% and 100% in turn.
3)
Check if the SpO2 value displayed on the screen is accurate. Error within 2% is
4)
acceptable.

10.2.5 NIBP Performance Test

This test checks the performance of NIBP measurement.
Required Equipment: Blood Pressure Simulator
Connect the equipment and the monitor as shown below:
1)
Blood Pressure
Switch on the monitor.
2)
Generate the NIBP signal 270/215 (235) and 40/10(20) in turn.
3)
Check if the NIBP value displayed on the screen is accurate. Error within 8 mmHg is
4)
acceptable.

10.2.6 TEMP Performance Test

This test checks the performance of TEMP measurement.
Required Equipment: Resistance Box
1)
Connect the equipment and the monitor as shown below:
measurement.
2
Simulator
2
SpO
Simulator
2
Simulator
- 62 -
NIBP Cuff
Periodic Tests