F15 Series Fetal & Maternal Monitor Service Manual
1
The monitor must be serviced by authorized and qualified personnel only. EDAN do
not accept responsibility for safety compliance, reliability and performance if
modifications or repairs are carried out by unauthorized personnel. Identical
replacement parts must be used.
2
The service personnel should be familiar with the operation of this monitor. Refer to
User Manual of the monitor for details.
3
EXPLOSION HAZARD - Do not use the monitor in the presence of flammable
anesthetics or other materials.
4
SHOCK HAZARD - the power receptacle must be a three-wire grounded outlet.
Never try to adapt the three-prong plug to fit a two-slot outlet. A hospital grade outlet
is required. If the outlet has only two slots, make sure that it is replaced with a
three-slot grounded outlet before attempting to operate the monitor.
5
Any non-medical equipment (such as the external printer) is not allowed to be used
within the patient vicinity (1.5m/6ft.).
6
Do not use the additional multiple portable socket-outlet or extension cord in the
medical electrical system, unless it's specified as part of the system by manufacturer.
And the multiple portable socket-outlets provided with the system shall only be used
for supplying power to equipment which is intended to form part of the system.
7
If multiple instruments are connected to a patient, the sum of the leakage currents
may exceed the limits given in the IEC/EN 60601-1 and may pose a safety hazard.
Consult your service personnel.
8
Multiple portable socket-outlets shall not be placed on the floor.
9
Make sure that the power is turned off and the power cord is disconnected from the
AC socket before connecting or disconnecting equipment. Otherwise, the patient or
operator may receive electrical shock or other injury.
10 Do not connect any equipment or accessories that are not approved by the
manufacturer or that are not IEC 60601-1 approved to the monitor. The operation or
use of non-approved equipment or accessories with the monitor is not tested or
supported, and monitor operation and safety are not guaranteed.
11 SHOCK HAZARD - Don't connect non-medical electrical equipment, which has been
supplied as a part of the system, directly to the wall outlet when the non-medical
equipment is intended to be supplied by a multiple portable socket-outlet with an
isolation transformer.
WARNING
9
Safety Guidance