F15 Series Fetal & Maternal Monitor Service Manual
4.14 Changing FHR Algorithm
The Beat-to-beat FHR algorithm provides vivid short-term variation information of the fetal
heart rate. The Average FHR algorithm provides a smoothened FHR trace.
1 Open the System Setup menu.
2 Select Beat-to-beat or Average.
3 Select OK.
4.15 Pressure Verification
Pressure Verification enables the Pressure Verif mode, in which you can verify whether there is a
problem with the NIBP measurement.
1
Open the System Setup menu.
2
Select Pressure Verif.
NOTE:
Make sure neither NIBP measuring nor NIBP leak test is being performed before you
enable pressure verification. If you need to disable pressure verification, please open the
System Setup menu and press Pressure Verif again. In addition, you should restart the
system after the pressure verification is finished.
4.16 NIBP Leak Test
NIBP Leak Test checks whether there is leak during NIBP measuring.
1
Open the System Setup menu.
2
Select NIBP Leak Test.
NOTE:
Make sure neither NIBP measuring nor pressure verification is being performed before
you enable NIBP leak test. Once the leak test is enabled, it can't be stopped manually. It
will automatically stop after it is finished and a message box will prompt out to inform you
whether the test has passed or not. In addition, you should restart the system after the
NIBP leak test is finished.
4.17 NIBP Calibration
NOTE:
1
NIBP calibration must be performed by professional personnel authorized by
EDAN.
2
NIBP calibration can influence measurement results. Incorrect operation may
influence measurement accuracy.
3
NIBP calibration can only be performed when none of NIBP measuring, pressure
verification and NIBP leak test is being performed.
Tools required:
- 17 -
System Configuration