System Check - GE CARESCAPE R860 Quick Reference Manual

Ventilator
Hide thumbs Also See for CARESCAPE R860:
Table of Contents

Advertisement

System Check

Circuit Setup
1. Select CIRCUIT SETUP.
2. Select check box for the HME or Humidifier.
Note:
Settings are confirmed when the setting
is changed.
3. Select X to close the menu.
Note:
Use Circuit Setup to select settings that must be
compensated for in the patient circuit measurements.
Running a System Check
1. From Standby, select SYSTEM CHECK.
Note:
The Run System Check menu shows.
2. Attach the breathing circuit and all accessories that
will be used to ventilate the patient.
Note:
Complete the system check using the appropriate
flow sensor per patient type. Use the neonatal flow sensor
when completing a system check for a neonatal patient.
3. Occlude the patient wye using the occlusion port.
Note:
The occlusion port sits below the exhalation valve
assembly (EVA).
4. Select Start.
Note:
When performing the resistance check, the wye-piece
and the breathing accessories (D-lite+
and HME) should remain on the occlusion port.
5. Select the information icon to see the System Check
Details menu.
a. As the system check is running, the results of each
check will be displayed.
Note:
Follow all on-screen system check instructions. A green
check mark will display if the check passes. If the check fails,
a red X is shown, and a Help icon displays next to the failed
check to provide possible causes and help for troubleshooting.
6
1
2
3
1
5
4
or Pedi-lite
Sensor
5a