Auto-Tuning(Pid Parameter Calibration) - Fisher Scientific Fisherbrand Isotemp RT AVCD HPS Advanced Operation Manual

Hotplate stirrer
Table of Contents

Advertisement

Check Configuration
Settings (LISt)
Auto-tuning (PID
Parameter Calibration)
Step 5: Terminate configuration mode as described in Configuration
Mode.
The PID parameters for temperature control can be automatically tuned
to take your specific operating circumstances into consideration.
If you select User Mode [Coefficient], the unit will be operated with the
auto tuning value.
The auto-tuning (calibration) procedures are as follows:
Step 1: Press the CONFIG Button.
Step 2: Select the auto-tuning mode [CALi] shown below by turning the
Stirrer Knob.
Step 3: Press the Stirrer Knob and select YES by turning the Stirrer
Knob. Press the Stirrer Knob again to start the auto-tuning
4-24
Section 4
Control Panel