8.4.7 Install Waste Module

After cleaning, flush the waste module with copious amounts of authorized rinsing agent (see Section 8.1) to remove any
traces of the disinfecting agent. Install new O-rings in the base plate under the waste module. Reinstall the waste module
using the three hex screws previously removed. Connect the end of the new calibrator tubing to the fitting on the side of
the waste module.

8.4.8 Waste Tubing

Slide the new waste tubing onto the large fitting on the side of the waste module.

8.4.9 Install Sample Module

After cleaning, flush the sample module with copious amounts of warm water, then rinse with authorized rinsing agent
(see Section 8.1) to remove any traces of the disinfecting agent. Install the sample module. Remember to install the stir
bar and the module seal O-ring. Secure the sample module using the three thumb nuts. Install a new O-ring on the
temperature probe, then install the temperature probe into the sample module.
Be sure to clean the enzyme probes before installing them in the sample module.

8.4.10 Sipper Replacement

Caution: Touch a bare metal chassis screw before handling the sipper tube
to prevent possible damage due to an electrostatic discharge.
The Sipper can be damaged if it is not properly aligned or if its alignment is disturbed. Inspect the Sipper for straightness
and condition of the Teflon jacket separating the two stainless steel tubes. If the Teflon jacket is torn, replacement of the
Sipper is required. Follow the steps below to replace the Sipper.
From the Service screen, Sipper tab, move the sipper to Location [Station 1-P96] to allow access to the sipper.
C1A
Calibrator and Waste Tubing
Figure 8-10
8-9
Connect waste
tubing to fitting