4 Basic Setup

The following list describes the basic steps necessary for sampling with the 2500 Analyzer.

1. Install Bottle Rack

2. Connect Printer (optional)
3. Connect AC Power
4. Align Sipper
5. Prepare and Install Buffer solution
6. Install Calibrator Solution
7. Prime the Fluid System
8. Install Membrane(s)
9. Configure Instrument Chemistries
10. Check Probe Currents
4.1
Install Bottle Rack
1. Install the Bottle Rack with Reagent Level Sensing onto the right side of the instrument by sliding the slots in the
tray over the pins on the side of the instrument.
2. Then remove the packing material holding the tubing to the right side of the instrument.
3. Next, connect bottle tubing and cables
a. Insert the large diameter waste tube into the hole in the waste bottle.
b. Connect one end of a short cable to the threaded fitting on the waste bottle cap and connect the other
end to the Waste (top) fitting on the instrument.
c. Connect the tubing marked "B1" and one end of a short cable to the fittings on the buffer bottle cap and
connect the other end of the cable to the Buffer (2nd) fitting on the instrument.
d. Connect the tubing marked "C1A" and one end of a long cable to the fittings on the calibrator bottle cap
and connect the other end of the cable to the CAL (bottom) fitting on the instrument.
Waste
Buffer
Not Used
CAL
Figure 4-1
4-1
Waste Bottle
Buffer Bottle
Calibrator Bottle
CAL