Service; Sipper - Xylem YSI 2500 User Manual

Biochemistry analyzer
Table of Contents

Advertisement

Touch the current date on the calendar to select it.
Touch the Hour button and enter the current hour in 24 hour format (0–23), then touch [OK].
Touch the Minute button and enter the current minute, then touch [OK].
When you have finished entering the date and time, touch [OK] to return to the main display.
6.2

Service

Touch [Service] to display the Service menu.

6.2.1 Sipper

See Section 4.4 Align Sipper for details on aligning the sipper with the sample module.
Always [Home] the sipper before performing alignment.
If necessary, the sipper can also be aligned with Station 2, the different types of racks/plates used at Station 1, and the
calibrator well.
6.2.1.1 Interlocks
From the Settings screen, System tab, touch the Interlocks button and change the status to [Off] to disable the interlock
switches on the front door and side panel when aligning the sipper.
WARNING: Keep your hands clear of the sipper while the instrument is in operation.
Always turn the Interlocks back [On] before operating the instrument!
6-20