Pro10
Quick Start Guide
This Quick Start Guide is meant to serve as a quick reference in calibrating and
operating the Pro10. It is not intended to replace the information found in the User
Manual.

Getting Started

Unpack the instrument and install (2) C size batteries in the back of the instrument.
Tighten the four screws of the battery cover on to the back of the instrument.

Installing the Sensor

Prior to use, install either a pH or ORP sensor into the 60510 cable assembly. Remove
the temporary plastic plug from the cable's port by pulling it straight out of the port.
Remove the plastic plug from the sensor's connector by pulling it straight off the sensor.
Insert the sensor into the port so the o-rings on the sensor are no longer visible. Hand
tightening the sensor into place by twisting it clockwise. Do not use a tool and do not
over tighten
Enable the correct ISE (either pH or ORP) in the instrument's System Setup menu. Attach
the cable assembly to your instrument.
pH Calibration
The Pro10 pH sensor can be calibrated by performing a 1, 2 or 3-point calibration. At
least one of the calibration points must be done with pH buffer 7 or 6.86. For highest
accuracy, perform at least a 2-point calibration.
2 or 3-Point Calibration
1.
Place the sensor in pH buffer 7 or 6.86 and allow the temperature and pH
readings to stabilize.
2.
Press and hold Cal for three seconds.
3.
Highlight pH and press Enter. If pH is not listed as an option, check the System
Setup menu to ensure pH is enabled in the ISE Sensor Type menu.
4.
Highlight 2 point or 3 point and press enter.
5.
If necessary, use the up and down arrow keys to adjust the pH buffer value.
6.
Press Enter to continue to second point.
Item #605193

Advertisement

loading

Summary of Contents for Xylem YSI Pro10

  • Page 1: Getting Started

    Pro10 Item #605193 Quick Start Guide This Quick Start Guide is meant to serve as a quick reference in calibrating and operating the Pro10. It is not intended to replace the information found in the User Manual. Getting Started Unpack the instrument and install (2) C size batteries in the back of the instrument. Tighten the four screws of the battery cover on to the back of the instrument.
  • Page 2: Orp Calibration

    Rinse the sensor and place it in the second pH buffer (4/4.01 or 10/9.18). If necessary, use the up and down arrow keys to adjust the pH buffer value. Wait approximately 30 to 60 seconds for the pH sensor to stabilize and for the temperature reading to stabilize.
  • Page 3: Taking Measurements

    Taking Measurements Before taking measurements, be sure the instrument has been calibrated to ensure the most accurate readings. Install the sensor guard and then insert the probe into the sample. Highlight Save and press Enter to store the reading. The instrument will confirm the reading was successfully saved.
  • Page 4 This page is intentionally left blank.