Preparation; Wiring - PCE Health and Fitness TDS 75 User Manual

Ultrasonic flow meter
Table of Contents

Advertisement

4

Preparation

4.1

Wiring

Open the device by loosening the four screws on the front to gain access to the ports for the
power supply, for the sensors, etc. First, lead the cables for the power supply through the cable
gland provided for this purpose and establish the power supply via the DC+ and DC- connections.
Ensure correct polarity. Then connect the inlet sensor and the outlet sensor to the corresponding
ports. For more information on the connection labels, refer to the following chart. When
connecting, make absolutely sure that there is no voltage.
Character
DC+
DC-
RL OUT+
RL OUT-
OCT OUT+
OCT OUT
GND
UP+
UP-
GND
DN+
DN-
I OUT+
I OUT-
TX
RX
GND
A
B
Attention!
Only wire the PCE-TDS 75 when it is switched off. The unit must be reliably grounded
before installation and use. Use either AC or DC power. Do not connect both at the
same time.
Description
Direct current DC 10 ... 36 V+
Direct current DC 10 ... 36 V
Grounding
Relay output
OCT output
Inlet sensor ground (black)
Inlet sensor + (brown)
Inlet sensor - (blue)
Outlet sensor ground (black)
Outlet sensor + (brown)
Outlet sensor - (blue)
4 ... 20 mA output
RS232 output
RS485 output