Step 5. Mounting method
Press "↓" to switch to the next submenu. Here,
you select 0.V, for example. Press ENTER to
confirm your selection.
Step 6. Sensor spacing
Open M14 and mount the sensors according to
the indicated distance and the selected
method.
Step 7 Display measured value
Open menu 01 to display the flow rate in m³/h.
Attention!
Always press the ENTER key first if you want to change a parameter. If "Change" is
still not possible after pressing the ENTER key, this means that the system is locked
by a password. To unlock, select "Unlock" in window M54 and enter the password
you have previously specified.