Levelflex FMP51, FMP52, FMP54
NOTICE
Poor grounding of the end of probe may cause measuring errors.
Apply a narrow sleeve which has good electrical contact to the probe.
NOTICE
Welding may damage the main electronics module.
Before welding: Ground the probe and dismount electronics.
6.3.3
Securing coax probes
For Ex-approvals: For probe lengths ³ 3 m (10 ft) a support is required.
Coax probes can be supported at any point of the outer tube.