The RED LED is OFF, indicating there is no data in memory
•
The
PRESS button is used to Start and Stop a Recording
Session
•
If the
PRESS button is not pressed for a period of one minute, the
instrument will enter a low power STANDBY mode and wait for
either another button press or the recording start time to arrive (if a
recording is scheduled). While in the low power STANDBY mode,
the LEDs will not blink.
•
A button press of 0.5 seconds will return the unit back to the normal
STANDBY mode
Event: Recording with Memory Cleared
When a recording starts, the logger will continue to record until one of the
following occurs:
•
The Session is complete
•
The Memory is full and the recording mode is Start/Stop
•
PRESS button is pushed until the STOP (ORANGE) LED
The
lights up and is released before the next LED lights
•
The Stop Recording command from the Simple Logger
Panel is sent to the unit
•
The battery voltage falls below 1.7V
Event: Recording with a Partial or Full Memory
If the RED LED is double-blinking, the memory must be cleared before
any further recording can be performed.
If the RED LED is single-blinking prior to starting a New Recording Ses-
sion, the memory is partially full.
To save, clear or check memory availability, use the Simple Logger
Control Panel software.
There may be instances where the GREEN LED is also double-blinking
indicating that the logger is still recording. The user can choose to stop
the Recording Session and download the saved session(s) and/or clear
the memory.
NOTE: The logger memory cannot be erased while in the Record mode. The
recording must be stopped first.