6.0 Installation

6.1. Device Placement

When selecting the right place for device, please consider following:
Put the device on smooth, horizontal and stable place.
Leave at least 10 cm of space around the device for adequate air circulation.
Leave enough space around the device, for easy access and maintenance.
Don't place the device where there are rapid temperature and humidity changes. Also avoid places where
the unit would be exposed to direct sunlight or next to devices that output large amounts of heat.
Avoid locations where the unit may be exposed to excessive shocks or vibrations.
Avoid locations subject to frequent power fluctuations or power losses.
Do not use the device in a flammable or explosive atmosphere.
NOTE: There has to be easy access to the shaker control panel and main plug in case of emergency.

6.2. Attaching the Power Cord

First connect the main power cord to the shaker then connect the plug end of the cord to a grounded
wall socket. Avoid lines on which powerful electric motors, refrigerators and similar devices are
connected.
Be careful when you plug the cord to a grounded wall socket.
Do not touch the plug with wet hands.
Do not pull the plug by the cord.

6.3. Starting Up Check List

Unpack and install the device as specified in the previous section.
Open the lid and attach the microtube platform or the microplate platform.
Close the lid.
Switch on the device using of the ON/OFF switch on the front panel The switch will illuminate to
indicate that power is on.
Check the running parameters and set new parameters if necessary. Please refer to the instructions
described in next chapters.

6.4. Platforms

There are two platforms available for VorTemp 1550™. The 50 mL Conical Tube Platform can accommodate
up to 8 x 50 mL test tubes. The 15 mL Conical Tube Platform accommodates up to 18 x 15 mL test tubes.
In order for the unit to perform correctly the platform needs to be balanced at all times. Tubes need to be
distributed evenly across the platform and they also need to be filled with equal amounts of liquid. Shaking
unbalanced loads will affect the performance of the VorTemp 1550 and may permanently damage the motor.
The following are examples of correct and incorrect tube distribution across the platform.
CORRECT
The platforms are easily interchangeable and don't require any tools. The platform should be inserted
onto four posts that fit into platform's corner rubber gaskets. To change the platform simply grab the
center piece of the platform and carefully pull straight up.
CORRECT
INCORRECT
INCORRECT