6- REMOVAL & REPLACEMENT
6.3 Fan Removal & Replacement
WARNING
!
UNPLUG THE UNIT FROM THE POWER SOURCE
BEFORE ATTEMPTING ANY REMOVAL OR
REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK.
A. Fan Removal
1. Separate Top from Bottom. Refer to 6.2, part A.
2. With #1 Phillips screwdriver, remove the 2 screws
securing the Fan. See Figure 6.5.
3. Remove Fan from the unit.
4. Remove Fan Baffle from fan. See Figure 6.5.
B. Installing Fan
1. Route Fan Harness underneath fan as shown in
Figure 6.6.
2. Install Fan Baffle onto fan as shown in Figure 6.6.
3. Place new Fan, Part Number 27266, into system
base.
4. Make certain the Fan Baffle sits onto System
Housing Baffle Tabs as shown in Figure 6.7.
5. Secure fan to housing with the 2 mounting screws.
See Figure 6.8.
NOTE:
Do not overtighten the screws. Overtightening will damage the
threads of the brass inserts.
6. Reassemble the Top and Bottom Housing. Refer to
6.2, part B.
Intelect
®
/Genisys
REMOVE SCREWS
ROUTE
HARNESS
AS SHOWN
16
®
Ultrasound Therapy Systems
FIGURE 6.5
FIGURE 6.6
BAFFLE
TABS
FIGURE 6.7
SECURELY TIGHTEN
MOUNTING SCREWS
FIGURE 6.8
FAN
BAFFLE
FAN
BAFFLE

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Intelect Genisys 2776 Intelect Genisys 2782 Intelect Genisys 2792 and is the answer not in the manual?

Questions and answers

Table of Contents