6- REMOVAL & REPLACEMENT
6.5 Ultrasound Board Removal & Replacement
!
WARNING
UNPLUG THE UNIT FROM THE POWER SOURCE
BEFORE
ATTEMPTING
REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK.
A. Ultrasound Board Removal
1. Separate Top from Bottom. Refer to 6.2, part A.
2. With #1 Phillips screwdriver remove the 4 screws
securing the Ultrasound Board to the unit. See
Figure 6.13.
3. Carefully remove Ultrasound Board from the
Control Board/Ultrasound Board Connector. See
Figure 6.14.
4. Remove Control Board/Ultrasound Board Connector
from the Control Board. Be careful not to bend or
damage any of the pins. See Figure 6.14.
B. Installing Ultrasound Board
1. Install Control Board/Ultrasound Board Connector
into new Ultrasound Board, Part Number 27269.
NOTE:
Install the connector with the long pins inserted into Ultrasound
Board. Make certain connector is completely seated against
Ultrasound Board. See Figure 6.15.
2. Place new Ultrasound board as shown in Figure
6.16 over Control Board and align connector pins up
with Control Board Connector.
3. Carefully press the connector and board into place
until the connector is completely seated into Control
Board.
4. Secure Ultrasound Board with the 4 retaining
screws.
NOTE:
Do not overtighten the screws. Overtightening will damage the
threads of the brass inserts.
5. Reassemble the Top and Bottom Housing. Refer to
6.2, part B.
ANY
REMOVAL
OR
Intelect
®
/Genisys
SCREW
SCREW
REMOVE US BOARD
FROM CONNECTOR
ALIGN CONNECTOR PINS
WITH CONTROL BOARD
CONNECTOR
18
®
Ultrasound Therapy Systems
FIGURE 6.13
REMOVE CONNECTOR
FROM CONTROL BOARD
FIGURE 6.14
INSTALL CONNECTOR TO
US BOARD AS SHOWN
FIGURE 6.15
FIGURE 6.16
SCREW
SCREW