5- TROUBLESHOOTING
5.1 Intelect/Genisys Software Error Messages (Continued)
CODE No.
TYPE MESSAGE
200
ERR_EEPROM
Error accessing the internal
EEPROM used to store
system configuration
settings and protocols.
201
ERR_APPL_SAVE_CAL_
DATA
Error attempting to save
data to applicator EEPROM
during applicator
calibration.
202
ERR_APPL_NOT_
CALLED_OK
General error attempting to
calibrate an applicator.
203
ERR_APPL_SAVE_
TIMEOUT
Error attempting to save
data to applicator EEPROM
during applicator calibration
PROBABLE CAUSES
1. Bad Control Board
1. Bad Applicator
2. Bad Applicator Cable
3. Applicator PCB Connector Bad
4. Bad Ultrasound Board
1. Bad Applicator
2. Bad Applicator Cable
3. Applicator PCB Connector Bad
4. Bad Ultrasound Board
1. Bad Applicator
2. Bad Applicator Cable
3. Applicator PCB Connector Bad
4. Bad Ultrasound Board
8
Intelect
®
/Genisys
®
Ultrasound Therapy Systems
POSSIBLE REMEDIES
A. Replace Control Board.
A. Replace with known good Applicator.
A. Replace with known good Applicator.
A. Replace Ultrasound Board.
A. Replace Ultrasound Board.
A. Replace with known good Applicator.
A. Replace with known good Applicator.
A. Replace Ultrasound Board.
A. Replace Ultrasound Board.
A. Replace with known good Applicator.
A. Replace with known good Applicator.
A. Replace Ultrasound Board.
A. Replace Ultrasound Board.