5- TROUBLESHOOTING

5.6 Ultrasound Tests

A. Equipment Required
1. Degassed Water. Refer to page 10 for Degassed
Water Recipes.
2. Ohmic Instruments UPM DT 10 or UPM DT 100
Ultrasound Power Meter.
3. Dissolved Oxygen Test Kit. Used to test oxygen
level of degassed water.
4. Known good Ultrasound Applicator

5.7 Ultrasound Applicator Identification Test

NOTE:
Use any Ultrasound Applicator for this test.
A. Ultrasound Applicator Identification Test
Procedures
1. Without Applicator installed, turn unit on.
2. Look at the display center. The Applicator not
detected icon should appear. See Figure 5.3.
3. Plug Applicator into Applicator port on the front
R.H. corner of the system. Applicator LED should
flash five (5) times. See Figure 5.4.
4. Look at the display center. The Applicator
detected icon should appear. See Figure 5.5.
B. Ultrasound Applicator Identification Test Results
1. Unit operates as described in steps 2 and 4.
a) Unit passed test.
2. "NO CAL" icon displays center of system display.
a) Applicator not calibrated.
b) Possible bad Applicator. Re-test with known
good Applicator.
c) Possible bad Ultrasound PCB.
3. Not detected icon displayed after ten seconds of
Applicator being plugged in.
a) Possible bad applicator. Re-test with known
good Applicator.
b) Possible bad Ultrasound Board.
®
Intelect
/Genisys
12
®
Ultrasound Therapy Systems
FIGURE 5.3
FIGURE 5.4
FIGURE 5.5