6- REMOVAL & REPLACEMENT

6.1 General

!
WARNING
UNPLUG THE UNIT FROM THE POWER SOURCE
AND/OR REMOVE BATTERY BEFORE ATTEMPTING
ANY REMOVAL OR REPLACEMENT PROCEDURES
TO PREVENT ELECTRICAL SHOCK.
A. Tools & Equipment Required
1. #1 Phillips Screwdriver
2. Flat Blade Screwdriver
6.2. Top & Bottom of Unit
A. Removing Top from Bottom
1. Place unit face down on a grounded work surface.
NOTE:
Use a clean, soft cloth or cushion underneath the display lens
to prevent scratching or damage to the lens.
2. Remove the plynth from the bottom of the system.
See Figure 6.1.
3. Remove the Mains Power Cord from the system.
See Figure 6.1.
4. Remove Battery Cover with the Flat Blade
Screwdriver and remove Battery from the system if
equipped. See Figure 6.2.
5. Remove the Rear Fan Grill from the system to gain
access to the two rear mounting screws. See Figure
6.2.
6. Remove the two front feet from the system to gain
access to the two front screws. See Figure 6.3.
7. Remove the four mounting screws from the bottom
housing using the #1 Phillips Screwdriver. See
Figure 6.3.
8. Carefully separate the upper and lower housings of
the system. It may be necessary to pry the housings
apart at the Applicator Hook. See Figure 6.4.
9. Disconnect the Power Supply and Fan Harnesses
from the Control PCB of the system. See Figure 6.4.
B. Installing Top to Bottom
1. With top laying face down, carefully place the bottom
housing in position over the top housing.
2. Connect the Power Supply and Fan Harnesses to
the Control PCB onto their respective connectors.
3. Make certain the front insert and the Applicator Hook
are in position, then lightly press housings together
to properly seat them.
4. Secure housings together with the four phillips head
screws.
NOTE:
Do not overtighten the screws. Overtightening will damage the
threads of the brass inserts.
5. Follow Steps 1 - 7 in 6.2, A above in reverse order to
complete reassembly of system.
Intelect
®
/Genisys
REMOVE
PLYNTH
REMOVE BATTERY
COVER AND BATTERY
REMOVE
FRONT FEET
CAREFULLY PRY APART
HERE IF NECESSARY
APPLICATOR
HOOK
15
®
Ultrasound Therapy Systems
REMOVE MAINS
POWER CORD
FIGURE 6.1
REMOVE
FAN GRILL
FIGURE 6.2
REMOVE
FOUR SCREWS
FIGURE 6.3
DISCONNECT POWER
SUPPLY AND FAN
HARNESSES
FIGURE 6.4