5- TROUBLESHOOTING
5.1 Intelect/Genisys Software Error Messages (Continued)
CODE No.
TYPE MESSAGE
300
CRIT_ERR_US_BOARD
Some type of critical
ultrasound board error has
occurred.
301
CRIT_ERR_US_BOARD_
NOT_DETECTED
No ultrasound board is
detected in unit.
302
CRIT_ERR_US_RESET
Ultrasound board has reset
for some unknown reason.
303
CRIT_ERR_US_BOARD_
READ
Error reading from
ultrasound board.
304
CRIT_ERR_US_
BOARD_WRITE
Error writing to ultrasound
board
305
CRIT_ERR_US_BOARD_
CALIBRATION_ERROR
Error calibrating applicator
PROBABLE CAUSES
1. Ultrasound Board to Control Board
Connecting Header loose or
disconnected.
2. Bad Ultrasound Board.
3. Bad Control Board.
1. Ultrasound Board to Control Board
Connecting Header loose or
disconnected.
2. Bad Ultrasound Board.
3. Bad Control Board.
1. Ultrasound Board to Control Board
Connecting Header loose or
disconnected.
2. Bad Ultrasound Board.
3. Bad Control Board.
1. Ultrasound Board to Control Board
Connecting Header loose or
disconnected.
2. Bad Ultrasound Board.
3. Bad Control Board.
1. Ultrasound Board to Control Board
Connecting Header loose or
disconnected.
2. Bad Ultrasound Board.
3. Bad Control Board.
1. Ultrasound Board to Control Board
Connecting Header loose or
disconnected.
2. Bad Ultrasound Board.
3. Bad Control Board.
9
Intelect
®
/Genisys
®
Ultrasound Therapy Systems
POSSIBLE REMEDIES
A. Reseat Connecting Header into
Control Board.
A. Replace Ultrasound Board.
A. Replace Control Board.
A. Reseat Connecting Header into
Control Board.
A. Replace Ultrasound Board.
A. Replace Control Board.
A. Reseat Connecting Header into
Control Board.
A. Replace Ultrasound Board.
A. Replace Control Board.
A. Reseat Connecting Header into
Control Board.
A. Replace Ultrasound Board.
A. Replace Control Board.
A. Reseat Connecting Header into
Control Board.
A. Replace Ultrasound Board.
A. Replace Control Board.
A. Reseat Connecting Header into
Control Board.
A. Replace Ultrasound Board.
A. Replace Control Board.