Linearity; Checking The Linearity; Adjusting The Linearity; Adjusting The Sensitivity - Sartorius Basic plus Service Manual

Hide thumbs Also See for Sartorius Basic plus:
Table of Contents

Advertisement

Note:
Important note:
Important note:
Note:

Linearity

You can only adjust the linearity in these balances using the Sartorius MC1
Server (version 4.4 and later).

Checking the Linearity

- Check the linearity for the entire weighing range in 5 steps.
- Compare the values displayed with the tolerance ranges given in the
"Adjustment/Calibration Data Sheets for Basic
- If the deviations exceed the tolerance ranges, the linearity must be
adjusted.

Adjusting the Linearity

You can only adjust the linearity in these balances using the Sartorius MC1
Server (version 4.4 and later).
- Activate the BPI mode (see page 16).
- Adjust the linearity using the Sartorius MC1 Server (version 4.4 and later).
(Please see the program description for the MC1 for instructions).
After working in the BPI mode, do not forget to set the write-protect so that the
Basic
balance returns to the standard data output mode (SBI Mode) and it is
plus
possible to attach peripherals.

Adjusting the Sensitivity

External Adjustment
There are two different procedures for this adjustment:
1. As described in this manual, or
2. Using the Sartorius MC1 Server (version 4.4 and later).
- Make sure that the menu code "1 9 1 - external adjustment/calibration
accessible" is set.
- Tare the balance by pressing
adjustment/calibration function.
- The calibration weight required appears in the display.
SARTORIUS Basic
Balances".
plus
; then press
BP-T-TA.WMF
plus
to initiate the
25