ZOLL Propaq M Service Manual page 86

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

Fluke Biomedical BP Pump/Bio-Tek Instruments BP Pump
Adult/Pediatric Mode
Simulator
Expected
Setting
Systolic
200/150
198
120/80
120
80/50
83
Neonate Mode
Simulator
Expected
Setting
Systolic
60/30
56
35/15
31
Fluke Biomedical BP Pump/Bio-Tek Instruments BP Pump2
Adult/Pediatric Mode
Simulator
Expected
Setting
Systolic
200/150
202
120/80
118
80/50
86
Neonate Mode
Simulator
Expected
Setting
Systolic
80/50
78
60/30
57
35/15
31
82
FastBP
Expected
Expected
Diastolic
Systolic
151
200
83
116
54
77
NeoBP
Expected
Diastolic
26
12
FastBP
Expected
Expected
Diastolic
Systolic
157
201
80
116
51
77
NeoBP
Expected
Diastolic
46
31
12
StepBP
Expected
Diastolic
149
84
53
StepBP
Expected
Diastolic
146
82
55
9650-000821-01 Rev. C