ZOLL Propaq M Service Manual page 27

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

SpO
Test (for SpO
2
Note: Perform this test if your unit has SpO
SpCO, and SpMet.
Fluke Index 2 SpO
Equipment
Disconnect any ECG cable or IBP probes from the Propaq M unit.
Test Setup
Do This...
1.
Power on the Propaq M unit without the
SpO
2.
Connect the
unit.
3.
Set the simulator to 70%
Attach the
4.
Set the simulator to 80%
5.
Set the simulator to 90%
6.
Disconnect the
9650-000821-01 Rev. C
feature without SpCO and SpMet)
2
Simulator (or equivalent), Masimo
2
sensor attached.
2
SpO
sensor to the Propaq M
2
SpO
2
SpO
sensor to the simulator.
2
SpO
2
SpO
2
SpO
simulator.
2
Propaq M Service Manual
only. See the following procedure to test SpO
2
®
Rainbow
Observe This...
Verify that the
displays INITIALIZING, then
goes blank.
Verify that
displays SEARCHING.
Verify that after a brief period of
and 60 PPM.
initialization, the
displays an
67% and 73%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
and 60 PPM. Verify that after a brief period of
initialization, the
displays an
77% and 83%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
and 60 PPM. Verify that after a brief period of
initialization, the
displays an
87% and 93%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
Maintenance Tests
®
SpO
finger probe
2
Pass/Fail
SpO
numeric
2
SpO
numeric
2
SpO
numeric
2
SpO
value between
2
o o
o o
SpO
numeric
2
SpO
value between
2
o o
o o
SpO
numeric
2
SpO
value between
2
o o
o o
,
2
23