Heart Rate Alarm Test
Fluke Impulse 7000DP Analyzer
Equipment
Do this...
1.
Press the Leads quick access key
(
select lead II.
2.
On the Propaq M unit:
3.
On the Propaq M unit, select
HR/PR Alarm in the Alarm Settings
menu and use the navigation keys
to do the following:
Use the navigation keys to exit the
Alarm Settings menu.
4.
On the Propaq M unit, press the
Alarm Cancel quick access key
(
5.
Remove a lead wire from the
simulator
6.
On the Propaq M unit, press
7.
Reattach ECG Lead wire to
simulator
8.
Set simulator to 30 BPM or lower.
9.
On the Propaq M unit, press
10.
Set the simulator to 120 BPM.
11.
On the Propaq M unit, press
9650-000821-01 Rev. C
)
I, II,
on the Propaq M unit and
III...
Press the More quick
access key (
) key to
access the second set of
quick access keys.
Press the Alarm quick
access key (
).
Press the Limits softkey.
Set the Upper Limit Enable
to ON.
Set the Lower Limit
Enable to ON.
Set the Lower limit to 35.
Set the Upper limit to 150.
).
.
.
Propaq M Service Manual
Observe this...
II displays next to the Leads quick
access key.
The Alarm Settings menu displays.
The red alarm suspension timer
displays at the top of the screen next
to the message area.
ECG
he ECG LEAD FAULT message
T
displays.
The Alarm Suspend key flashes
.
ECG
The ECG LEAD FAULT message no
longer displays.
Heart Rate Value is highlighted and
the alarm tone sounds. The HR/PR
LOW ALARM message displays.
The Alarm Suspend key flashes.
.
red alarm suspension timer displays
at the top of the screen next to the
message area.
Alarm is enabled. The alarm suspend
.
key flashes. The alarm suspension
timer disappears.
Maintenance Tests
Pass/Fail
o o
o o
o o
.
o o
The
o o
o o
17