Temperature Test
IBP Temperature Simulator (or equivalent)
Equipment
Do this...
1.
Connect the temperature simulator to the
Propaq M T1 channel.
2.
On the IBP Temp Simulator, Select 98.6
Degree F (36 Degree C) or a setting as close
as possible as your simulator allows.
3.
Repeat steps 1 and 2 of this procedure for the
T2 channel.
4.
Disconnect cable.
9650-000821-01 Rev. C
Observe This...
The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.
Propaq M Service Manual
Maintenance Tests
Pass/Fail
o o