ZOLL Propaq M Service Manual page 25

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

Leakage Current Test
See the manufacturer's instructions or supplied specifications for the leakage tester you use.
Equipment
Test Setup
See the manufacturer's instructions or supplied specifications for the leakage tester you use.
Repeat leakage test with accessories: MFC, external paddles, internal paddles, and anterior/
posterior paddles. Perform these tests at the line-power voltage and frequency used in your
installation.
Maximum Leakage Acceptance Limits
ECG
MFC
Earth Leakage
9650-000821-01 Rev. C
Table 1:
Patient Leakage
Normal
Single Fault
10
50
10
50
Table 2:
Normal
2500
Propaq M Service Manual
Mains on Applied Part
Single Fault
100
100
Single Fault
5000
Maintenance Tests
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents