ZOLL Propaq M Service Manual page 37

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

NIBP Calibration Check
NIBP Simulator, single or dual lumen hose
Equipment
Do This...
1.
Turn on the Propaq M unit.
2.
Connect a single lumen hose from the
NIBP simulator to the NIBP port of the
Propaq M.
3.
On the Propaq M unit, do the following to
enter the NIBP Test control panel:
Note: You need a supervisor passcode to
enter the Supervisor menu.
4.
From the Propaq M NIBP Test control
panel, use the navigation keys to select:
5.
Monitor the pressure reported by the
external pressure standard.
6.
Once the pressure reaches the target
pressure of 50 +/- 5 mmHg, use the
navigation keys to select Turn Pump Off.
7.
Allow the system to stabilize for at least
10 seconds.
8.
Verify that the primary transducer agrees
with the reading on the NIBP simulator.
9.
Verify that the safety transducer agrees
with the reading on the NIBP simulator.
10. From the NIBP Test control panel, select
Open Valve to release the pressure.
11. From the Propaq M NIBP Test control
panel, use the navigation keys to select:
12. Monitor the pressure reported by the
external pressure standard.
9650-000821-01 Rev. C
Press the More (
) then the Setup
(
) quick access keys.
Use the navigation keys to select
Supervisor>Service>Device
Tests>NIBP Test.
Close Valve
Turn Pump On
Close Valve
Turn Pump On
Propaq M Service Manual
Observe This...
Reading should be within 3mmHg.
Reading should be within 4mmHg.
Maintenance Tests
Pass/Fail
o o
o o
33

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents