ZOLL Propaq M Service Manual page 33

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

EtCO
Calibration (required for annual calibration)
2
5% CO
Equipment
2
Products, PN 4653ORF-BD)
Do This...
Turn on the Propaq M unit.
1.
2.
On the Propaq M unit, select
Supervisor>Service>Diagnostic>CO2>
Calibration to enter the CO
control panel.
3.
Connect the Propaq M unit to the 5% CO
calibration gas cylinder using the sample
line included with the calibration kit.
4.
From the CO
Calibration.
5.
Activate the calibration gas flow by
pressing the top valve on the calibration gas
canister.
6.
Continue to apply the calibration gas until the
status in the CO
displays: CALIBRATION GAS CAN BE
REMOVED.
The calibration date will be updated upon
exiting the CO
9650-000821-01 Rev. C
calibration gas cylinder with CO
Calibration Menu, select
2
Calibration control panel
2
When the calibration is complete, the
status in the CO
Calibration control panel
2
displays: CALIBRATION DONE
Test control panel.
2
Propaq M Service Manual
sample line (kit available from Scott Medical
2
Observe This...
Calibration
2
2
Maintenance Tests
Pass/Fail
o o
o o
29