Device Cleaning And Disinfecting; Cuff Cleaning And Disinfecting; Accuracy Test; Disposal - Microlife WatchBP Office Vascular Instruction Manual

Professional office blood pressure and cardiovascular screening monitor
Hide thumbs Also See for WatchBP Office Vascular:
Table of Contents

Advertisement

Device cleaning and disinfecting

Use a soft cloth with one of the following recommended
cleaning solutions to wipe the exterior of the device:
• Ethyl or isopropyl alcohol (70% solution).
• Hydrogen peroxide 7.5% solution.
• Sodium hypochlorite solution (5.25-6.15%
household bleach diluted 1:500 provides >100 ppm
available chlorine)
Then wipe the exterior of the device with a soft, dry cloth.
*
Do not immerse or spray the device.

Cuff cleaning and disinfecting

Take out the bladder. Fold and place the cuff cover
inside a washing bag. Wash the cuff cover with warm
water (43°C; 110°F) and a mild detergent in the
washing machine.
Pasteurization: wash the cuff cover in 75°C(167°F)
hot water for 30 minutes.
Warning:
• Under no circumstances may you wash the cuff bladder.
• The bladder must lay flat inside the cuff cover without folding.
Caution:
• Do not iron the cuff.
• Please disinfect before the next person uses it.

Accuracy test

We recommend the device to be tested for accuracy every 2 years or after mechanical
impact (e.g. Being dropped). Please contact Microlife to arrange an accuracy test.
Warning: The device and accessories are only serviced (tested & calibrated) by trained personnel of the Microlife
distributor or manufacturer. Do not attempt to service or calibrate the device and accessories yourself.

Disposal

Batteries and electronic instruments must be disposed of in accordance with the
locally applicable regulations, and not as domestic waste.
WatchBP Office
MODE
19