Endress+Hauser Liquiline To Go Ex CYM291 Operating Instructions Manual page 51

Table of Contents

Advertisement

Specifications
Memosens pH input
(also ISFET)
Display ranges
1)
Memosens input
ORP
Display ranges
1)
Sensor standardization *
Permissible calibration range
Sensor standardization *
Operating modes *
"Programmed buffers"
buffer sets *
Permissible calibration range
Calibration timer *
Sensoface
Evaluation of
* User-defined
1) Ranges depending on Memosens sensor
M8 socket, 4 pins, for Memosens lab cable or
M12 socket, 8 pins, for Memosens sensors
pH
mV
Temperature
M8 socket, 4 pins, for Memosens lab cable or
M12 socket, 8 pins, for Memosens sensors
mV
Temperature
ORP calibration (zero adjustment)
ΔmV (offset)
pH calibration
PROG.BUFFERS
MANUAL
DATA INPUT
-01- Endress+Hauser
-02- Mettler-Toledo
-03- Knick CaliMat
-04- Ciba (94)
-05- NIST technical
-06- NIST standard
-07- HACH
-08- WTW techn. buffers 2.00/4.01/7.00/10.00
-09- Hamilton
-10- Reagecon
-11- DIN 19267
Zero point
With ISFET:
Operating point
(asymmetry)
Slope
(possibly restricting notes from Sensoface)
Interval 1 ... 99 days, can be switched off
Provides information on the sensor condition
zero/slope, response, calibration interval
-2.00 ... +16.00
-1999 ... +1999 mV
-50 ... +250 °C (-58 ... +482 °F)
-1999 ... +1999 mV
-50 ... +250 °C (-58 ... +482 °F)
-700 ... +700 mV
Calibration with automatic buffer recognition
Manual calibration with entry of individual
buffer values
Data entry of zero and slope
2.00/4.01/6.98/9.95/11.87
2.00/4.01/7.00/9.21
2.00/4.00/7.00/9.00/12.00
2.06/4.00/7.00/10.00
1.68/4.00/7.00/10.01/12.46
1.679/4.006/6.865/9.180
4.01/7.00/10.01/12.00
2.00/4.01/7.00/10.01/12.00
2.00/4.00/7.00/9.00/12.00
1.09/4.65/6.79/9.23/12.75
6 ... 8 pH
-750 ... +750 mV
Approx. 74 ... 104 %
51
pH

Advertisement

Table of Contents
loading

Table of Contents