Specifications
Memosens pH input
(also ISFET)
Display ranges
1)
Memosens input
ORP
Display ranges
1)
Sensor standardization *
Permissible calibration range
Sensor standardization *
Operating modes *
"Programmed buffers"
buffer sets *
Permissible calibration range
Calibration timer *
Sensoface
Evaluation of
* User-defined
1) Ranges depending on Memosens sensor
M8 socket, 4 pins, for Memosens lab cable or
M12 socket, 8 pins, for Memosens sensors
pH
mV
Temperature
M8 socket, 4 pins, for Memosens lab cable or
M12 socket, 8 pins, for Memosens sensors
mV
Temperature
ORP calibration (zero adjustment)
ΔmV (offset)
pH calibration
PROG.BUFFERS
MANUAL
DATA INPUT
-01- Endress+Hauser
-02- Mettler-Toledo
-03- Knick CaliMat
-04- Ciba (94)
-05- NIST technical
-06- NIST standard
-07- HACH
-08- WTW techn. buffers 2.00/4.01/7.00/10.00
-09- Hamilton
-10- Reagecon
-11- DIN 19267
Zero point
With ISFET:
Operating point
(asymmetry)
Slope
(possibly restricting notes from Sensoface)
Interval 1 ... 99 days, can be switched off
Provides information on the sensor condition
zero/slope, response, calibration interval
-2.00 ... +16.00
-1999 ... +1999 mV
-50 ... +250 °C (-58 ... +482 °F)
-1999 ... +1999 mV
-50 ... +250 °C (-58 ... +482 °F)
-700 ... +700 mV
Calibration with automatic buffer recognition
Manual calibration with entry of individual
buffer values
Data entry of zero and slope
2.00/4.01/6.98/9.95/11.87
2.00/4.01/7.00/9.21
2.00/4.00/7.00/9.00/12.00
2.06/4.00/7.00/10.00
1.68/4.00/7.00/10.01/12.46
1.679/4.006/6.865/9.180
4.01/7.00/10.01/12.00
2.00/4.01/7.00/10.01/12.00
2.00/4.00/7.00/9.00/12.00
1.09/4.65/6.79/9.23/12.75
6 ... 8 pH
-750 ... +750 mV
Approx. 74 ... 104 %
51
pH