Configuring

Configuration (pH)

Configuration is required to match the connected sensor and the
desired measurement performance. Furthermore, you can select the
suitable calibration method. The following table gives you an over-
view. Factory settings are shown in bold print.
Measurement
"Setup" display
Display 1
Display 2
CAL Timer
CAL
CAL POINTS
BUFFER SET
(PROG.BUF-
FERS, FREE
CAL)
Auto OFF
Temp Unit
Time Format
Date Format
Default
This icon prompts you to select a menu item using the arrow keys –
the selection is confirmed by pressing .
Select using arrow keys, confirm by pressing .
pH x.xx | pH x.xxx | mV
OFF | date + time | date | time
OFF | 1 ... 99 days
PROG.BUFFERS|Manual|DATA INPUT|(ISFET-Zero)|
FREE CAL
1 | 2 | 3 | 1-2-3 (for PROG.BUFFERS, Manual, FREE CAL)
-01- Endress+Hauser
-02- Mettler-Toledo
-03- Knick CaliMat
-04- Ciba (94)
-05- NIST technical
-06- NIST standard
-07- HACH
-08- WTW techn. buffers 2.00|4.01|7.00|10.00
-09- Hamilton
-10- Reagecon
-11- DIN 19267
OFF| 0.1h | 1h | 6h | 12h
°C | °F
24h | 12h
dd.mm.yy | mm.dd.yy
NO | YES (reset to factory settings)
Note: All data logger entries will be deleted.
2.00|4.01|6.98|9.95|11.87
2.00|4.01|7.00|9.21
2.00|4.00|7.00|9.00|12.00
2.06|4.00|7.00|10.00
1.68|4.00|7.00|10.01|12.46
1.679|4.006|6.865|9.180
4.01|7.00|10.01|12.00
2.00|4.01|7.00|10.01|12.00
2.00|4.00|7.00|9.00|12.00
1.09|4.65|6.79|9.23|12.75
17