Endress+Hauser Liquiline To Go Ex CYM291 Operating Instructions Manual page 25

Table of Contents

Advertisement

Conductivity Calibration
Calibration with Calibration Solution
(Automatic calibration with preselected calibration solution)
The calibration method is selected in the configuration menu.
NOTICE!
• Make sure that the values of the calibration solutions used
correspond exactly to those specified in this manual.
If not, the resulting cell constant will be incorrect.
• When calibrating in a liquid, make sure that the sensor, the
separate temperature probe (if present) and the calibration
solution have the same temperature. Only this ensures that the
cell constant is determined correctly.
Measurement
CAL
<25°C> 107.00mS/cm
PRESS CAL
Measured value
Temperature
Cal solution conductivity
Hourglass blinks
Calibration will be performed. Automatic return to measuring mode.
Dip sensor in calibration solution.
The meter automatically compensates for the
temperature deviation!
25