10.4
Replacing power end bearings
8.
As soon as the system back pressure of the system has been
reached, determine the fill level Q
9.
Run the pump for a while.
10.
Simultaneously determine the fill level Q
11.
Calculate the dosing rate value.
12.
Reduce the dosing rate using the stroke adjustment dial and repeat
steps 7 to 10 until the required value is reached.
13.
Turn the indicating dial (1) only until the 100% marking is precisely
above the indicating slot (2).
14.
Carefully tighten the screw (3) in the stroke adjustment dial and
replace the cap (4).
15.
API versions only: Press the metal cap on to the stroke adjustment
dial.
Only allow ProMinent Service to replace the power end bearings!
Repair
and start the stopwatch.
1
and stop the stopwatch.
2