Installing Two Filters In The Filter Holder - FLIR X6520sc Series User Manual

Table of Contents

Advertisement

7
Operation
4. Place your filter in the holder. Take care to ensure correct filter orientation, to avoid er-
rors in the radiometric measurement. Contact your filter provider for this information.
5. Gently insert the threaded filter ring, and tighten it using the filter tool. Take great care
not to damage the filter with the tool.

7.1.5 Installing two filters in the filter holder

You can install two filters in the filter holder.
Note
• The total thickness of the two filters must not exceed 2 mm.
• The order in which you install the filters will not affect their performance. However, to
avoid narcissus effects, it is recommended that the filters are installed with the least
reflective side downwards in the filter holder (i.e., toward the detector).
• Filters are fragile. Handle them with the great care.
• Do not touch the filter surface when you install the filters. If this happens, clean the fil-
ter according to the manufacturer's instructions.
• Wear gloves or finger tips to handle the filters
• A set of filter holder equipment (filter holder, centering ring, filter spacer, threaded filter
ring, and magnets) and a filter holder tool are provided with the camera.
Follow the procedure below to install two filters in the filter holder.
1. Create a custom filter ID, as described in section 7.1.4 Creating a custom filter holder,
page 29.
2. Place the first filter in the holder.
#T810207; r. AA/43061/43086; en-US