PNA Series Microwave Network Analyzers
N5241A/42A

Rear Panel Troubleshooting

Each rear panel connector is associated with a hardware group in the analyzer. You can
use the data at these rear panel connectors to help troubleshoot these hardware groups in
addition to testing the connectors.
The connectors discussed in this section are:
• USB x 4
• Monitor (VGA)
• GPIB (0) CONTROLLER
• GPIB (1) TALKER/LISTENER
• LAN

Checking the USB Ports

To verify proper operation of any rear panel USB port:
• Connect a known good USB device, such as a USB mouse.
• Wait 15 seconds for the analyzer to verify the device connection, and then check the
operation of the USB device.
• If the device performs correctly, the USB port is functioning properly.
• If the device does not perform correctly, remove the non-working USB device, wait 15
seconds, and then reconnect the device to the rear panel USB port.
• If the USB device still does not work and has been verified to work elsewhere, then the
A17 CPU board is faulty. Refer to
Assembly" on page

Checking the CONTROLLER Port

The network analyzer uses a National Instruments 488.2 GPIB controller and associated
driver software. This software includes a test utility that scans the GPIB bus and returns
the status of all the connected peripherals.
To run the test utility software and check the GPIB status:
1. Connect a known good peripheral to the analyzer using a known good GPIB cable.
2. Press UTILITY
System
box is displayed.
3. In the
block, click
GPIB
for the analyzer to configure, and then click
4. If the Window Desktop is not displayed, press UTILITY
Control Panel...
to view the
Service Guide N5242-90001
"Removing and Replacing the A17 CPU Board
7-22.
, then
Configure
System Controller
Windows Taskbar
, then
SICL/GPIB...
. A
to establish the analyzer as a controller. Wait
.
OK
System
menu at the bottom of the display.
Troubleshooting
Rear Panel Troubleshooting
dialog
SICL/GPIB/SCPI
, then
Configure
, then
4-17