Repair and Replacement Procedures
PNA Series Microwave Network Analyzers

Removing and Replacing the A16 Power Supply Assembly

N5241A/42A
Removing and Replacing the A16 Power Supply Assembly
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-9
for this procedure.
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to
"Removing the Covers" on page
7-6.
3. Remove the three flat head screws (item
) from the power supply bracket.
4. Remove the seven pan head screws (item
) from the power supply rear panel.
5. Slide the A16 power supply assembly out the rear of the analyzer.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
Table 7-2 on page
7-68.
7-20
Service Guide N5242-90001

Advertisement