PNA Series Microwave Network Analyzers
N5241A/42A

Power Supply Check

There are no fuses to replace within the power supply. If you determine that
NOTE
the power supply is the failed assembly, replace the power supply.
A catastrophic failure in the power supply can be determined by observing the line switch
and the power supply LED indicators:
1. Ensure that the instrument is plugged in with the power switch in the standby position
(power not switched on). Verify that the line switch displays a yellow light. A line switch
that displays a yellow light indicates that the power supply standby line is active and
functional)
2. Turn on the instrument power and verify that the line switch now displays a green
light. When the line switch displays a green light, it is an indication that the power
supply is active and does not sense an over-current condition.
3. You can determine which power supplies are functioning by viewing the LED indicators
on the A15 midplane board. Refer to
To view the LED indicators, it is necessary to remove the instrument's outer and inner
covers. Refer to
"Removing the Covers" on page 7-6
determine the location of the A15 midplane board, refer to
Options" on page
CAUTION
Do not operate the analyzer with the outer cover removed for more than
30 minutes, as this could cause the analyzer to overheat which could result in
costly damage.
4. If any power supply voltage is missing, it is likely that the problem is a defective A16
power supply, the A15 midplane board, or another assembly that is loading down the
A16 power supply. Continue with
determine the cause of the problem.
5. If the line switch is lit correctly, and all the power supply voltages appear to be present,
as indicated by the LEDs as shown in
catastrophic failure. However, the power supply could still be at fault. Continue at
"Measure the Individual Supply Voltages"
correct.
Measure the Individual Supply Voltages
Measure the power supply voltages using a digital multi-meter. Use the point labeled
ACOM as ground reference for analog supplies and the point marked DCOM as ground
reference for digital supplies.
Refer to
Figure 4-2
for the power supply measurement points on the A15 midplane board.
Refer to
Table 4-1 on page 4-9
Service Guide N5242-90001
Figure
6-14.
"If Any Supply Voltage Is Missing" on page 4-9
for the correct voltages.
4-2.
for removal procedures. To
Figure
4-2, the power supply has not suffered a
to verify that the actual supply voltages are
Troubleshooting
Power Up Troubleshooting
"Top Assemblies, All
to
4-7