Frequency Accuracy Test; Trace Noise Test - Agilent Technologies N5242A Service Manual

2-port and 4-port pna-x microwave network analyzers
Hide thumbs Also See for N5242A:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
N5241A/42A

Frequency Accuracy Test

Function of the Test: To verify the frequency accuracy and range of the analyzer's source
output.
Specification Tested: Test Port Output–CW Accuracy
Equipment Used: A frequency counter, a test cable, and adapters.
Description of the Test:
1. Port 1 is connected to a frequency counter.
2. A series of frequencies across the band are checked.
If the Analyzer Fails this Test:
• Verify the accuracy of the 10 MHz OCXO by using a frequency counter to measure the
rear-panel 10 MHz REF OUT. If the 10 MHz reference is off by more than 10 Hz,
perform the
"10 MHz Frequency Reference Adjustment" on page 3-37
this test.

Trace Noise Test

Function of the Test: To measure the stability of a signal in the internal source and
receiver system of your analyzer.
Specification Tested: Test Port Input–Trace Noise Magnitude and Trace Noise Phase
Equipment Used: A test cable.
Description of the Test:
1. Port 1 and Port 2 are connected with a test cable.
2. The analyzer is set to a series of CW frequencies across its frequency range.
3. Magnitude and phase are measured at each frequency at both 1 kHz and 10 kHz IF
bandwidths, in both directions.
4. Measurements are made at a nominal power level of 0 dBm and 201 points per sweep.
If the Analyzer Fails this Test:
• Repeat this test. It is unlikely for the analyzer to fail this test without a complete
failure of the network analyzer system.
• If the analyzer still fails this test, replace the A12 SPAM board and then repeat this
test. Refer to
"Removing and Replacing the A4–A13 Boards" on page
Service Guide N5242-90001
Tests and Adjustments
Performance Tests
and then repeat
7-14.