PNA Series Microwave Network Analyzers
General Purpose Maintenance Procedures
N5241A/42A

Correction Constants

Correction Constants
The analyzer stores many correction constants in non-volatile EEPROM memory. These
constants enable the analyzer to produce accurate, leveled source signals and receive clean
test signals.

Storing Correction Constants

After performing any adjustment listed on
page 3-36
in this manual, store the correction
constants to a backup file on the analyzer hard disk drive by performing these steps:
• Navigate to the EEPROM Backup Utility, located at:
C:\Program Files\Agilent\Network Analyzer\Service\eebackup.exe
• Run the program.
• Click Backup EEPROM.
• Click Exit when the program has finished.
Service Guide N5242-90001
8-19