Measurement Traceability - Agilent Technologies N5242A Service Manual

2-port and 4-port pna-x microwave network analyzers
Hide thumbs Also See for N5242A:
Table of Contents

Advertisement

Tests and Adjustments
PNA Series Microwave Network Analyzers
System Verification
N5241A/42A

Measurement Traceability

To establish a measurement traceability path to a national standard for a network
analyzer system, the overall system performance is verified through the measurement of
devices that have a traceable path. This is accomplished by measuring the devices in an
Agilent verification kit.
The measurement of the devices in the verification kit has a traceable path because the
factory system that measured the devices is calibrated and verified by measuring
standards that have a traceable path to the National Institute of Standards and
Technology (NIST) (see
Figure
3-8). This chain of measurements defines how the
verification process brings traceability to the network analyzer system.
Figure 3-8
NIST Traceability Path for Calibration and Verification Standard
3-22