Removing And Replacing The Front Panel Assembly - Agilent Technologies N5242A Service Manual

2-port and 4-port pna-x microwave network analyzers
Hide thumbs Also See for N5242A:
Table of Contents

Advertisement

Repair and Replacement Procedures

Removing and Replacing the Front Panel Assembly

Removing and Replacing the Front Panel Assembly
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16 inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-2
for this procedure.
1. Disconnect the power cord.
2. Remove the outer cover. Refer to
3. With a 5/16 inch torque wrench, remove all the semirigid jumpers (item
front panel.
4. With a T-10 TORX driver, remove the 12 screws (item
CAUTION
Before removing the front panel from the analyzer, lift and support the front
of the analyzer frame.
5. Slide the front panel over the test port connectors.
6. Disconnect the ribbon cable (item
Replacement Procedure
IMPORTANT
When reconnecting the front-panel jumpers, torque the connectors to 10 in-lb.
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
7-8
"Removing the Covers" on page
) from the A1 front panel interface board.
Table 7-2 on page
7-68.
PNA Series Microwave Network Analyzers
7-6.
) from the
) from the sides of the frame.
Service Guide N5242-90001