Repair and Replacement Procedures

Removing and Replacing the Lithium Battery

Removing and Replacing the Lithium Battery
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• ESD grounding wrist strap
Removal Procedure (2.0 GHz CPU Board)
Refer to
Figure 7-30
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to
3. Remove the A17 CPU board. Refer to
Assembly" on page
4. Remove top cover attachment screws (item
5. Lift the top cover off of the CPU board assembly.
6. Remove the battery from the battery holder by lifting it at the open end of the holder
and then sliding it from under the clip (item
7. DO NOT THROW AWAY THE BATTERY. COLLECT IT AS SMALL CHEMICAL
WASTE. Refer to
battery disposal.
7-64
for this procedure.
"Removing and Replacing the A17 CPU Board
7-22.
"Lithium Battery Disposal" on page 1-8
PNA Series Microwave Network Analyzers
"Removing the Covers" on page
) and (item
).
).
for additional information on
Service Guide N5242-90001
N5241A/42A
7-6.