PNA Series Microwave Network Analyzers
N5241A/42A
Source Adjustment
The source calibration is used to adjust your network analyzer for a flat source power
across its full frequency range. There are differences between each test port; therefore, an
adjustment is required for each port.
Equipment Used for the Source Adjustment
Equipment Type
Power meter
Power sensor, 3.5 mm
Adapter, 3.5 mm (f) to 3.5 mm (f)
Procedure
1. Connect the equipment as shown in
network analyzer and the power meter.
Figure 3-14 Equipment Setup for the Source Adjustment
2. .Press UTILITY
System
Adjustment.
3. Ensure the GPIB settings are correct.
4. Follow the instructions and prompts as they are displayed.
Service Guide N5242-90001
Figure
3-14. Connect a GPIB cable between the
, then
Service
, then
Model or
Part Number
E4418B/E4419B
E4413A
83059B
Adjustments
, then click Source
Tests and Adjustments
Adjustments
Alternate Model or
Part Number
E4418A/E4419A
8485A
85052-60012
3-39