NOTE! 
Check the hygiene cover each time the product is cleaned. If damaged, it must be 
 
replaced or repaired. Also check the pump, tube connectors and power cord during 
 
cleaning. Damaged parts must be replaced or repaired. 
 
 
 
 
6. Storage 
 
The mattress and pump are best stored in the product bag (optional extra). Handle the packed 
product with care. Do not put anything heavy on top of it. 
 
 
7. Maintenance
 
7.1 General 
 
1. Check the power cord and plug for damage and scratches. 
 
2. Check that the mattress cover is intact and that the cover and air cells are correctly 
assembled. 
 
3. Check that the air flow alternates between each outlet on the side of the pump. 
 
4. Check for damaged or pinched tubes and connectors. 
 
Contact the manufacturer or your local dealer for spare parts. 
 
7.2 Changing fuses  
 
 
 
7.3 Changing the air filter  
 
 
 
 
1.   P ull the plug out of the electrical socket before replacing fuses.  
 
2.   R emove the cover for the fuses with a screwdriver. Pull out 
   the fuse holder.  
 
3.   F it a new fuse of the same type (T1AL, 250 V) and screw  
the protective cover back in place. There are two fuses. 
 
 
1. Loosen the small round cover plate at the back of the pump. 
 
2. Lean the pump back and the filter slides out of its holder. 
 
3. Place the new filter in the holder and put back the cover plate.  
The filter should be cleaned regularly if the pump is used in a 
very dirty environment. 