5.2 Reconditioning
PUMP
See section 5.1 for instructions.
MATTRESS
1. Disconnect the tubes and let the air out, also open the CPR valve.
2. Remove the cover. See instructions in chapter 5.1 for cleaning covers. If the product is
equipped with a foam core this should be removed and cleaned separately.
3. Move the mattress to a clean area, which is suitable for cleaning.
4. Wipe off the cells, all tubes and the CPR valve with a mild cleaning solution for this purpose
such as: alcohol with or without tensides, Isopropanol 70 %, oxidating solutions or chlorine
max 1 %.
5. Wipe off and disinfect all the work surfaces with an alcohol-based cleaning agent intended
for this purpose.
6. Put the mattress together by putting back the cell package in the cover and place the pump
tubes by the hole on the short side.
7. Fasten the cell package in its holder in all corners. Adjust and close the zipper.
NOTE!
Check the hygiene cover each time the product is cleaned. If damaged, it must be
replaced or repaired. Also check the pump, tube connectors and power cord during
cleaning. Damaged parts must be replaced or repaired.

6. Storage

The mattress and pump are best stored in the product bag (optional extra). Handle the packed
product with care. Do not put anything heavy on top of it.