Advertisement

5. Reuse, cleaning and reconditioning

The product is reusable. When reusing, it is important to
follow the instructions below for cleaning and
reconditioning. Before using the product again for a new
user, it is recommended to clean the mattress in accordance
with the instructions below. Consult your hygiene manager
or Care of Sweden for help and instructions if you are unsure.
Cleaning and reconditioning should otherwise be performed
as required.
5.1 Cleaning
CONTROL UNIT
Wipe off the control unit with a damp cloth and mild
cleaning solution for this purpose: such as alcohol
with or without tenside surfactants, oxidizing
solutions, chlorine max 1 % or hydrogen peroxide max 1,5 %.
If using a different agent, choose one that does not affect the
control unit.
MATTRESS COVER
Covers can be wiped off with a mild cleaning solu‐
tion for this purpose such as: alcohol with or with‐
out tenside surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % (max 10% for cover Stone) or
hydrogen peroxide max 1,5 %.
Excrement and blood stains must be removed as soon as
possible using cold water. Carefully follow local instructions
and the instructions for the detergent.
Multi‐layer covers should be divided before washing. The
parts should be washed with similar colours.
Chlorine and phenol‐based cleaners could adversely
affect the PU surface and should be avoided. If chlo‐
rine is used, we recommend a mixture of max 1 %
(max 10% for cover Stone).
INNER COVER
Clean the affected area with a mild cleaning solution for this
purpose such as: alcohol with or without tenside surfactants,
Isopropanol 70 %, oxidizing solutions, chlorine max 1 % or
hydrogen peroxide max 1,5 %.

6. Storage

It is advisable to store the mattress and control unit in the
product bag (accessory). Handle the packaged product with
care. Do not place any heavy objects on top of it. For addi‐
tional information about storage temperature, see section 9.
5.2 Reconditioning
CONTROL UNIT
See section 5.1 for instructions.
MATTRESS
1. Disconnect the tube connector from the control unit and
remove the air from the mattress.
2. Remove the covers. For more information of cleaning the
covers, see section 5.1.
3. Move the mattress to a clean area that is suitable for
cleaning.
4. Wipe off the cells, all tubes and the CPR module with a
mild cleaning solution for this purpose such as: alcohol with
or without tenside surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % or hydrogen peroxide max 1,5 %.
Let it dry.
5. Wipe off the work surfaces using a suitable detergent and/
or disinfectant.
6. Put the mattress together. If cells for any reason have be‐
come loose from the tubes, these must be replaced accord‐
ing to the drawing in section 2.1.
NOTE:
Check the hygiene cover, cells and hoses each time the prod‐
uct is cleaned. If damaged, it must be replaced or repaired.
Also check the control unit, tube connectors and power cable
during cleaning. Damaged parts must be replaced or repaired.
The air cells can be divided using a quick‐release connector,
making it easier to clean the cells.
If the control unit has a backup battery (variant) this must be
charged up every six months if the control unit is not in use.
The control unit should be stored in a dry environment
indoors and out of direct sunlight.