3 - 4
Replacement and wiring of PS/2 cable
The PS/2 cable in the monitor trolley is discarded and replaced by the PS/2 cable supplied
with the cable set.
•
Remove the old PS/2 cable completely.
To do this, cut open the cable ties and remove them.
•
Run the new PS/2 cable from the PS/2 plug of the PC (imaging system) parallel to the
existing cables of the CIPP board (D66) to the ON/OFF assembly D50 and connect it to
connector D50.X13.
•
Attach the new cable to the existing ones with cable ties. Make sure there is no
mechanical strain (tension on the cable) on the PS/2 connector.
Wiring of adapter cable D50.X12- X22
•
Wire up the adapter cable D50.X12 - X22 supplied with the cable set.
•
To to this, connect the plug connection designated X22 with connector X22
and plug the other end D50.X12 into the ON/OFF assembly.
Wiring on D50
•
Plug in connector D50.X9 and screw it on. Make sure that connector D50.X16 is not bent
or disconnected.
•
Plug in connector D50.X8.
•
Plug in connector D50.X6.
•
Plug in connector D50.X7.
•
Plug in connector D50.X13.
ARCADIS
System Manual
Installing the ON/OFF assembly
n.a.
SP00-000.841.01
Rev. 01
Page 4 of 10
08.05
CS PS 24
3
3
3
Siemens AG
Medical Solutions